Download-button_s

Experiment Overview

Repository ID: FR-FCM-ZY2E Experiment name: Data aquisition of different input light intensities (quantiFlash) to PMT voltages from 400-1000V to determine laser background, signal-to-noise and dynamic range of FC channels MIFlowCyt score: 57.57%
Primary researcher: Toralf Kaiser PI/manager: Toralf Kaiser Uploaded by: Toralf Kaiser
Experiment dates: 2016-12-01 - 2017-01-31 Dataset uploaded: Feb 2017 Last updated: Sep 2017
Keywords: [quantiFlash] [instrument characterization] [dynamic range] [signal-to-noise] Manuscripts: [28960720] Cytalogo
Organizations: Deutsches Rheuma-Forschungszentrum Berlin, Flow Cytometry Core Facility (FCCF), Berlin, (Germany)
Purpose: Data aquisition of different input light intensities (quantiFlash) to PMT voltages ranging from 400-1000V (and 50V increment steps) to determine laser background, signal-to-noise and dynamic range of FC channels, i.e. instrument characterization using a stable LED light source
Conclusion: a precision input-light source (quantiFlash) can be used to determine instruments background (by switching the lasers off), and SNR and DNR can be calculated from background to a reference light-input signal
Comments: None
Funding: Kristen Feher's contributions were made possible by European Union project 10158448EFRE. Claudia Giesecke's contributions were made possible by DFG funding through the Berlin-Brandenburg School for Regenerative Therapies GSC 203.
Quality control: we repeated these experiemnts and on different instruments and came to the same conslusions
Download FCS Files or login and see the dataset in your inbox for further annotation details.