Repository ID: FR-FCM-ZZRJ Experiment name: Dilution series 40nm SA-AgNP on CD8-bio MIFlowCyt score: 77.25%
Primary researcher: Axel Schulz PI/manager: Axel Schulz Uploaded by: Axel Schulz
Experiment dates: 2016-03-01 - 2016-03-02 Dataset uploaded: May 2016 Last updated: Mar 2019
Keywords: [nanoparticle] [Titration] [mass cytometry] Manuscripts: [27351740]
Organizations: German Rheumatism Research Center Berlin, a Leibniz-Institute (DRFZ), Immunemonitoring, Berlin, (Germany)
Purpose: Determine optimal concentration for usage of 40 nm SA-AgNP in mass cytometry Done on PBMC with CD8-bio as primary antibody Repetition of an older experiment for publication
Conclusion: Best signal with conc 1:25, but problems in all other channels (incomplete ionization, due to high metal content? (1:25 - 1:100). Best signal without this issue: 1:200. No sufficient resolution at titers above 1:400
Comments: see Fig 1 D-E
Funding: Not disclosed
Quality control: Comparison to direct detection by CD8-168Er (serves also as no primary ab control)
Download FCS Files or login and see the dataset in your inbox for further annotation details.